What does status code 00211 mean on the LCD display of DC 500 and 800 series spectro
Status code ’00211’ is an indicator that the green tile result is high. This is an indicator that the instrument has a problem or if it is a loaner or recently serviced instrument the green tile data needs to be deleted and reset.